Tech Tip 13: The importance of the use of the correct liner for the sample introduction technique selected

In the Sample Introduction section of our Complete GC & GC-MS course we highlight the importance of selecting the correct liner style for the injection type that you are going to perform. Some of the questions to ask are:

  • Is it a solvent, gas or a solventless injection that you are performing and what vapour volume will it produce if any? This will affect the volume of the liner.
  • Are you trying to get all of the sample onto the column or are you going to flush most out of the split exit? This will affect the style of the liner.
  • Are you trying to efficiently mix the sample with the carrier gas, wipe the end of the needle for viscous samples or trying to trap dirt? This will affect the position of any glass wool present.
  • Are your analytes active and low concentration? This will affect the degree of deactivation and if to use glass wool at all!

Selecting the wrong liner for the sample introduction technique can have disastrous consequences including poor sensitivity, poor reproducibility, lost peaks, contamination of the inlet and gas lines, carryover, ghost peaks, poor peak shapes and the list continues.

One example is trying to use a large internal diameter (ID) liner (4 mm) for a solventless Solid-Phase Micro-Extraction (SPME) injection in splitless mode. With a SPME injection the objective is desorb the fiber in the inlet and quickly transfer the analytes onto the analytical column in a narrow sample band to get sharp peaks. Long desorption or transfer times allow the volatile analyte molecules to migrate down the analytical column and produce broad peaks, which have a much lower signal to noise as well as poor resolution. A large ID liner has a large volume which takes longer to flush, resulting in the broad peaks shown in the chromatogram. Using the recommended, much narrower 0.75 mm ID SPME liner helps both the desorption but mainly the transfer of analytes onto the column, as it takes much less time to perform a complete flush of the liner, giving sharp Gaussian peaks, with a vastly improved S/N and baseline resolution.

The resulting chromatograms are shown here; the chromatogram with the nice sharp peaks is as a result of correct liner use, whilst the one with the broad peaks were due to the incorrect liner.

There are many liners on the market, if in doubt about the type of liner to use, ask yourself the above questions which are the main points to think about.

To learn more about sample introduction techniques for GC & GC-MS, attend Day 1 / Module 2 of our Complete GC & GC-MS course.

Don't forget that you can also subscribe to our newsletters and receive useful tips straight to your inbox!